14.1 Changing the Fuse
Make sure that the mains cable is unplugged before changing the fuse. This operation should
only be carried out by personnel familiar with the transmitter and who are qualified for such
work.
If the power consumption of the M400 transmitter is too high or a manipulation leads to a
short circuit the fuse will blow. In this case remove the fuse and replace it with one specified
in Section 15.
14.2 Warning- and Alarm list pH
Warnings Description
Warning pH slope > 101% Slope too big
Warning pH Slope < 95% Slope too small
Warning pH Zero > 7.5 pH Zero offset too big
Warning pH Zero < 6.5 pH Zero offset too small
Warning pHGls change < 0.3 Glass electrode resitance changed by more than factor 0.3
Warning pHGls change > 3 Glass electrode resitance changed by more than factor 3
Warning pHRef change < 0.3 Reference electrode resitance changed by more than factor 0.3
Warning pHRef change > 3 Reference electrode resitance changed by more than factor 3
Alarms Description
Watchdog time-out SW/System fault
Error pH Slope > 102% Slope too big
Error pH Slope < 90% Slope too small
Error pH Zero > 8.0 pH Zero offset too big
Error pH Zero < 6.0 pH Zero offset too small
Error pH Ref Res >150 KΩ Reference electrode resistance too big (break)
Error pH Ref Res < 2000 Ω Reference electrode resistance too small (short)
Error pH Gls Res > 2000 MΩ Glass electrode resistance too big (break)
Error pH Gls Res < 5 MΩ Glass electrode resistance too small (short)
14.3 Warning- and Alarm list O
2
Warnings Description
Warning O
2
Slope < –90nA Slope too big
Warning O
2
Slope > –35nA Slope too small
Warning O
2
ZeroPt > 0.3nA Zero offset too big
Warning O
2
ZeroPt < –0.3nA Zero offset too small
Alarms Description
Watchdog time-out SW/System fault
Error O
2
Slope < –110 nA Slope too big
Error O
2
Slope > –30 nA Slope too small
Error O
2
ZeroPt > 0.6 nA Zero offset too big
Error O
2
ZeroPt < –0.6 nA Zero offset too small