Transmitter Multiparameter M400 69
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
12.3 Model / Software Revision
Selecting Model / Software Revision will display the model number, serial number and installed
software revision.
The displayed information is important for any Service call. Press [ENTER] to return to the
normal measurement mode.
12.4 ISM Sensor Info (available when ISM sensor connected)
After plugging in an ISM sensor, the following information about the sensor will be shown in this
menu. Use up and down arrows to scroll in the menu.
Type: Type of sensor (e.g. InPro 3250)
Cal Date: Date of the last adjustment
Serial-No.: Serial number of the connected Sensor
Part-No.: Part number of the connected Sensor
12.5 ISM Sensor Diagnostics (available when ISM sensor connected)
Cal History
The calibration history is stored with a time stamp in the ISM Sensor and is displayed on the
transmitter. The Calibration history offers the following information:
Fact (Factory calibration): This is the original dataset, determined in the factory. This dataset
remains stored in the sensor for reference and cannot be overwritten.
Act (Actual adjustment): This is the actual calibration dataset which is used for the
measurement. This dataset moves to Cal-2 position after the next adjustment.
1. Adj (First adjustment): This is the first adjustment after the factory calibration. This dataset
remains stored in the sensor for reference and cannot be overwritten
Cal-1 (last calibration / adjustment): This is the last executed calibration / adjustment. This
dataset moves to Cal-2 and then to Cal-3 when a new calibration / adjustment is performed.
Afterwards, the dataset is not available anymore.
Cal-2 and Cal-3 acting in the same way as Cal-1.