ISM (digital) Sensor
If an ISM sensor is connected, the transmitter automatically (Parameter = Auto) recognizes
the type of sensor. You can also fix the transmitter to a certain measurement parameter
(Parameter = pH / ORP, Cond (2), Cond(4), O
2
hi or 0
2
lo), depending on the type of transmitter
you have. Choose the display setting for line a, b, c, d.
Measurement parameter Type
pH / ORP = pH or ORP 1,2,3
Cond (2) = 2 electrode conductivity* 1,2,3
Cond (4) = 4 electrode conductivity* 1,2,3
O
2
hi = Dissolved oxygen (ppm) 2,3
O
2
lo = Dissolved oxygen (ppb, traces*) 3
O
2
opt hi = Dissolved oxygen (ppm) optical 2,3
O
2
opt lo = Dissolved oxygen (ppb) optical 3
* Available as of Q4 / 2008
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values and return to the measurement display screen, selecting Yes will
save changes made.
8.2.2 Temperature Source (Not used with ISM Sensors)
Press the [ENTER] key to select this: Menu / Configure / Measurement / Temperature Source.
The following options can be chosen:
Auto: The transmitter automatically recognizes the temperature source.
Use NTC22K: Input will be taken from the sensor attached.
Use Pt1000: Temperature input will be taken from the sensor attached.
Use Pt100: Input will be taken from the sensor attached.
Fixed = 25 °C: Allows a specific temperature value to be entered. It must be chosen when
customer use pH sensor without temperature source.
8.2.3 pH / O
2
Parameter related settings
Press [ENTER] to select this: Menu / Quick Setup / Channel select ISM / Parameter.
Additional measurement and calibration parameters can be set for each parameter; conductivity,
pH and O
2
.
For more details, please see the following explanations depending on the selected parameter.
8.2.3.1 Conductivity Temperature Compensation
The temperature compensation mode for any of the four measurement lines can be selected.
Temperature compensation should be matched to the characteristics of the application.
Choices are ”Standard” and ”Linear”. Press [ENTER] and save changes.
Standard compensation includes compensation for non-linear high purity effects as well as
conventional neutral salt impurities and conforms to ASTM standards D1125 and D5391.