Transmitter Multiparameter M400 32
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
7 Sensor Calibration
(PATH: Cal)
The calibration key allows the user one-touch access to Sensor calibration and verification
features.
C
NOTE: During Calibration on Channel A, a flashing ”H” (Hold) in the upper left corner of the
display indicates a calibration is in process with a Hold condition active. (The hold output
function needs to be activated.) See also chapter 3.3.
7.1 Enter Calibration Mode
While in Measurement mode press the key.
Press the or key to select the type of calibration desired.
After selecting ”Sensor”, use the key to move to the next line. Select the channel ”A” or ”B” to
be calibrated. Select the desired Sensor Calibration task. The choices for each sensor type are:
Conductivity = Conductivity, Resistivity, Temperature*, Edit*, Verify
Oxygen = Oxygen, Temperature*, Edit*, Verify
pH = pH, mV, Temperature*, Edit pH*, Edit mV, Verify
Press [ENTER].
* only on channel ”A”
After every successful calibration, the three options are available:
Adjust: Calibration values will be overtaken und used for the measurement. Additionally,
the data will be stored in the calibration history*.
Calibrate: Calibration values will be stored in the calibration history* for documentation,
but will not be used for the measurement. The calibration values from the last valid
Adjustment will be further used for the measurement.
Abort: Calibration values will be discarded.
* only available with ISM sensors
7.2 Conductivity Calibration
This feature provides the ability to perform a one-point or two-point Conductivity ”Sensor”
calibration. The procedure described below works for both types of calibrations. There
is no reason to perform a two-point calibration on a two-electrode conductivity sensor. Four
electrode sensors do require a two-point calibration.
C
NOTE: When performing calibration on a conductivity sensor, results will vary depending on the
methods, calibration apparatus and / or quality of reference standards used to perform the
calibration.
Enter Conductivity Sensor Calibration mode as described in section 7.1.
Choose ”standard” or ”linear” compensation mode.
Press [ENTER].