11.2.1.2 Current
Current Calibration is preformed as a two point calibration.
Navigate to the Calibrate Meter screen and select Channel A.
Enter the value for Point 1, in milliamps, of the current source connected to the input. The
second display line will show the measured current. Press
[ENTER] to begin the calibration process.
Enter the value for Point 2, in milliamps, of the current source connected to the input. The
second display line shows the measured current.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes
to save the calibration values and the successful Calibration is confirmed on the display. The
transmitter will return to the Measurement mode in approximately 5 seconds.
11.2.1.3 Voltage
Voltage Calibration is preformed as a two point calibration.
Navigate to the Calibrate Meter screen and select Channel A and Voltage.
Enter the value for Point 1 in, volts, connected to the input. The second display line will show
the measured voltage. Press[ENTER] to begin the calibration process.
Enter the value for Point 2, in volts, of the source connected to the input. The second display
line shows the measured voltage.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes
to save the calibration values and the successful Calibration is confirmed on the display. The
transmitter will return to the Measurement mode in approximately 5 seconds.