Mettler Toledo M400 Especificações Página 62

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 61
Transmitter Multiparameter M400 62
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
11.1.3 Display
All pixels of the display will be lit for 15 seconds to allow troubleshooting of the display. After
15 seconds the transmitter will return to the normal Measuring mode or press [ENTER] to exit
sooner.
11.1.4 Keypad
For keypad diagnostics, the display will indicate which key is pressed. Pressing [ENTER] will
return the transmitter to the normal Measuring mode.
11.1.5 Memory
If Memory is selected then the transmitter will perform a RAM and ROM memory test. Test
patterns will be written to and read from all RAM memory locations. The ROM checksum will
be recalculated and compared to the value stored in the ROM.
11.1.6 Set Relay
The Set Relays diagnostic menu allows for the activation / deactivation of eachRelay. To access
relays 5 and 6, press [ENTER].
0 = Normal (normally open contacts are open)
1 = Inverted (normally open contacts are closed)
Press [ENTER] to return to Measurement mode.
Vista de página 61