Transmitter Multiparameter M400 37
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
Z is the newly calculated offset calibration factor. The slope calibration factor S is always 1
and does not enter the calculation.
After a successful calibration, the calibration values are taken over and stored in the cal history
(Adjust), only stored in the cal history (Calibrate) or aborted.
7.5 Sensor Temperature Calibration
Enter Sensor Calibration mode as described in section 7.1 and select Temperature.
7.5.1 One-Point Sensor Temperature Calibration
Select 1 point calibration. Slope or Offset can be selected with the 1 point calibration. Select
Slope to recalculate the Slope factor M (Multiplier) or Offset to recalculate the offset calibration
factor A (Adder).
Enter the value for Point 1 and press [ENTER].
After a successful calibration, the calibration values are taken over and stored in the cal history
(Adjust), only stored in the cal history (Calibrate) or aborted.
7.5.2 Two – Point Sensor Temperature Calibration
Select 2 point as calibration Type.
Enter the value for Point 1 and press [ENTER].
Enter the value for Point 2 and press [ENTER].