Mettler Toledo M400/2(X)H Especificações Página 91

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 90
Transmitter M400/2(X)H, M400G/2XH 91
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400/2(X)H, M400G/2XH
Printed in Switzerland 30 031 683
14.4.3 Trace oxygen sensors
Warnings Description
Warning O
2
Slope < 5000 nA Slope too big
Warning O
2
Slope > 3000 nA Slope too small
Warning O
2
ZeroPt > 0.5 nA Zero offset too big
Warning O
2
ZeroPt < 0.5 nA Zero offset too small
Alarms Description
Watchdog time-out SW/System fault
Error O
2
Slope < 6000 nA Slope too big
Error O
2
Slope > 2000 nA Slope too small
Error O
2
ZeroPt > 1.0 nA Zero offset too big
Error O
2
ZeroPt < 1.0 nA Zero offset too small
Electrolyte Low* Too low level of electrolyte
* ISM sensors only