The Autoclaving Cycle Limit counts the number of autoclaving cycles. If the limit (user defined)
is reached, an alarm can be indicated and set to a certain output OC. The following options are
possible:
Autoclave Max 000
Alarm YES / NO R# choose OC
If the Max setting is on 000, the counter functionality is turned off. The alarm will be reset after
exchanging the sensor. For oxygen sensors, the counter can also be reset manually (see
chapter “Reset ISM counter/timer”).
Autoclave characteristics: Since during the autoclaving cycle the sensor is not connected to the
transmitter, you will be asked after every sensor connection, whether the sensor was autoclaved
or not. According to your selection, the counter will be incremented or not.
Pressing the [ENTER] key will bring up the Save Changes dialog. Selecting No will discard the
entered values, selecting Yes will make the entered values the current ones.
8.6.5 Reset ISM counter / timer
This menu allows resetting counter and timer functions which cannot be reseted automatically.
The adaptive calibration timer will be reseted after a successful adjustment or calibration.
Navigate to the menu “Reset ISM Counter/Timer” by using the and . keys and press
[ENTER].
If an pH sensor or amperometric oxygen sensor is connected, the menu for resetting the Time To
Maintenance is displayed. Time To Maintenance needs to be reset after the following operations.
pH sensors: manual maintenance cycle on the sensor.
oxygen sensor: manual maintenance cycle on the sensor or exchanging of the inner-body or
membrane of the sensor
[Press ENTER]
If an oxygen sensor is connected, the menu for resetting the CIP and SIP counter is displayed.
These counters should be reset after the following operations.
amperometric sensor: exchanging of the inner-body of the sensor.
[Press ENTER]
8.6.6 DLI stress adjustment (only for pH ISM sensors)
Through this menu the calculation of the diagnostic data DLI, TTM and ACT can be adapted to
application requirements and/or experience.
h
NOTE: The function is only available for pH ISM sensors with corresponding firmware versions.