19 Buffer tables
M400 transmitters have the ability to do automatic pH buffer recognition. The following tables
show different standard buffers that are automatically recognized.
19.1 Standard pH buffers
19.1.1 Mettler-9
Temp (°C) pH of buffer solutions
0 2.03 4.01 7.12 9.52
5 2.02 4.01 7.09 9.45
10 2.01 4.00 7.06 9.38
15 2.00 4.00 7.04 9.32
20 2.00 4.00 7.02 9.26
25 2.00 4.01 7.00 9.21
30 1.99 4.01 6.99 9.16
35 1.99 4.02 6.98 9.11
40 1.98 4.03 6.97 9.06
45 1.98 4.04 6.97 9.03
50 1.98 4.06 6.97 8.99
55 1.98 4.08 6.98 8.96
60 1.98 4.10 6.98 8.93
65 1.98 4.13 6.99 8.90
70 1.99 4.16 7.00 8.88
75 1.99 4.19 7.02 8.85
80 2.00 4.22 7.04 8.83
85 2.00 4.26 7.06 8.81
90 2.00 4.30 7.09 8.79
95 2.00 4.35 7.12 8.77