Transmitter M400/2(X)H, M400G/2XH 83
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400/2(X)H, M400G/2XH
Printed in Switzerland 30 031 683
12.3 Model / Software revision
(PATH: Info/Model/Software Revision)
Selecting Model/Software Revision will display the part number, model and the serial number of
the transmitter.
By using the . key it is possible to navigate forward through this menu and get additional infor-
mation like the current version of firmware implemented on the transmitter (Master V_XXXX and
Comm V_XXXX) and – if an ISM sensor is connected – the version of the sensor firmware (Sen-
sor FW V_XXX) and sensor hardware (Sensor HW XXXX).
The displayed information is important for any Service call. Press [ENTER] to exit from this
display.
12.4 ISM sensor info
(available when ISM sensor connected)
(PATH: Info / ISM Sensor Info)
After plugging in an ISM sensor it is possible by using the key A or . to navigate to the Menu
“ISM Sensor Info”.
Press [ENTER] to select the menu.
The following information about the sensor will be shown in this menu. Use up and down
arrows to scroll in the menu. Type: Type of sensor (e.g. InPro 3250)
Cal Date: Date of the last adjustment
Serial-No.: Serial number of the connected sensor
Part-No.: Part number of the connected sensor
Press [ENTER] to exit from this display.
12.5 ISM sensor diagnostics
(available when ISM sensor connected)
(PATH: Info / ISM Diagnostics)
After plugging in an ISM sensor it is possible by using the key A or . to navigate to the Menu
“ISM Diagnostics”.
Press [ENTER] to select the menu.
Navigate to one of the menus, described in this section, and press [ENTER] again.
Comentários a estes Manuais