Enter the value for Point 1 of the calibration according to the resistor connected across the pH
glass electrode measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the pH
glass electrode measuring input.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes to
save the calibration values and the successful calibration is confirmed on the display. The trans-
mitter will return to the measurement mode in approximately 5 seconds.
11.2.1.5 Rr diagnostic
Rr diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen and
select Channel A and Rr Diagnostic.
Enter the value for Point 1 of the calibration according to the resistor connected across the pH
reference measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the pH
reference measuring input.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes to
save the calibration values and the successful calibration is confirmed on the display. The trans-
mitter will return to the measurement mode in approximately 5 seconds.
11.2.1.6 Calibrate analog output signals
Select the Analog Output you wish to calibrate. Each analog output can be calibrated at 4 and
20 mA.