Transmitter M300 83
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
The transmitter will return to the Measurement mode in approximately 5 seconds.
11.3.1.3 Current
Current Calibration is preformed as a two point calibration.
Navigate to the Calibrate Meter screen and select Channel A or B and Current.
Enter the value for Point 1, in milliamps, of the current source connected to the input. The sec-
ond display line will show the measured current. Press
[ENTER] to begin the calibration process.
Enter the value for Point 2, in milliamps, of the current source connected to the input. The sec-
ond display line shows the measured current.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes
to save the calibration values and the successful Calibration is confirmed on the display. The
transmitter will return to the Measurement mode in approximately 5 seconds.
11.3.1.4 Voltage
Voltage Calibration is preformed as a two point calibration.
Navigate to the Calibrate Meter screen and select Channel A or B and Voltage.
Enter the value for Point 1 in, volts, connected to the input. The second display line will show
the measured voltage. Press[ENTER] to begin the calibration process.
Enter the value for Point 2, in volts, of the source connected to the input. The second display
line shows the measured voltage.