The Lockout – Menu is protected: Enter the Administrator Password and select YES to enable or
NO to disable the Lockout functionality. Pressing the [ENTER] key after the selection will bring
up the Save Changes dialog. Selecting No will discard the entered value, selecting Yes will make
the entered value the current one.
9.5 Reset
(PATH: Menu/System/Reset)
This Menu allows access to the following options:
Reset System, Reset Meter Cal*, Reset Analog Cal, Reset Cal Data**.
* Not on ISM version
** For ISM version of Cond 2-e sensors only.
9.5.1 ResetSystem
This Menu allows the reset of the meter to the factory default settings (Setpoints off, analog out-
puts off, etc.). The meter calibration and the analog output calibration are not affected.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the me-
ter.
9.5.2 ResetMeterCalibration(notonISMversion)
This Menu allows the reset of the meter”s calibration factors to the last factory calibration values.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the meter
calibration factors.