Navigate to the Calibrate Meter screen and select Channel A or B, and Resistance 1, designating
that the transmitter is ready to calibrate the first range resistor. This resistance may be changed,
selecting range 1 thru 5. Each resistance range consists of a two-point calibration.
Press [ENTER] to begin the calibration process.
The first text line will ask for the Point 1 resistance value (this will correspond to Resistance 1
value shown on the Calibration Module Accessory). The second text line will show the measured
resistance value. When the value stabilizes, press [ENTER] to perform calibration.
The transmitter screen will then prompt the user to enter the value for Point 2, and R1 will dis-
play the measured resistance value. When this value stabilizes, press [ENTER] to calibrate this
range and bring up a confirmation screen.
Select Yes to save the calibration values and the successful Calibration is confirmed on the dis-
play. The transmitter will return to the Measurement mode in approximately 5 seconds.
Once point 1 and 2 are calibrated, return to the Calibrate Meter screen. Move the cursor to
change to Resistance 2, designating the second calibration range. Proceed with the two-point
calibration process as performed for the first range. This same process must be followed to
complete the resistance calibration of all 5 ranges.
11.3.1.2 Temperature
Temperature is performed as a three point calibration. The table above shows the resistance val-
ues of these three points.
Navigate to the Calibrate Meter screen and choose Temperature calibration for either Channel A
or B.
Press [ENTER] to begin temperature calibration process
The first text line will ask for the Point 1 temperature resistance value (this will correspond to
Temperature 1 value shown on the Calibration Module Accessory). The second text line will
show the measured resistance value. When the value stabilizes, press [ENTER] to perform cali-
bration.
The transmitter screen will then prompt the user to enter the value for Point 2, and T2 will
display the measured resistance value. When this value stabilizes, press [ENTER] to calibrate
this range.
Repeat these steps for Point 3.
Press [ENTER] to bring up a confirmation screen. Select Yes to save the calibration values and
the successful Calibration is confirmed on the display.