Transmitter M300 72
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
9.5.3 ResetAnalogCalibration
This Menu allows reset of the Analog Output calibration factors to the last factory calibration val-
ues.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the Ana-
log Output calibration.
9.5.4 ResetSensorCalibrationDatatoFactorySettings
If a Cond 2-e sensor based on ISM technology is connected to the transmitter, this menu is
available. The menu allows the reset of the calibration data (M resp. A) of the sensors to the fac-
tory settings.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the cali-
bration data of the sensor to factory settings.
h
NOTE: To ensure best measuring results, a new calibration of the sensor is recommended after a
reset of the calibration data to factory settings. Depending on the application, the calibration
could be temporarily performed as a process calibration, recommended however is an one point
calibration (see chapter 7.2 “Conductivity/Resistivity Calibration”).
9.5.5 ResetCalibrationDataofSensorElectronictoFactory
Settings
If a Cond 2-e sensor based on ISM technology is connected to the transmitter, this menu is
available. The menu allows the reset of the calibration data of the sensor`s evaluation electronic
to the factory settings.
The procedure for this functions is described in the sensor manual.