Transmitter M300 FLOW 51
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M300 FLOW
Printed in Switzerland 52 121 319
10.1 Enter PID Setup
While in Measurement mode press the key. Press the or key to navigate to the PID
Setup – Menu and press [ENTER]. Enter the System security code ”xxxxx” if required, (see
section 9.3) and press the [ENTER] key to confirm the code.
C
Note: to exit PID Setup mode at any time press the and keys simultaneously
(Escape). The transmitter returns to the Measurement mode and the old values
remain active.
10.2 PID Display Auto-Manual
(PATH: Menu/PID Setup/PID A/M Manual)
This menu allows selection of Automatic or Manual operation. Select Auto or Manual operation.
Pressing the [ENTER] key will bring up the Save Changes dialog.
10.3 Tune Parameters
(PATH: Menu/PID Setup/Tune Parameters)
This menu assigns control to a measurement and sets the tuning parameters, setpoint,
and non-linear functions of the controller.
10.3.1 PID Assignment & Tuning
Assign the measurement, a, b, c, or d (single channel models) or A, B, C, D, a, b, c, d
(four channel transmitters) to be controlled. Set the Gain (unitless), Integral or Reset time,
Tr (minutes) and Rate or Derivative time, Td (minutes) needed for control. Press [ENTER].
Gain, Tr and Td are later adjusted by trial and error based on process response. Td is usually
left at zero for flow control.
10.3.2 Setpoint & Deadband
Enter the setpoint value and the deadband around the setpoint, where no proportional control
action will take place (normally zero for flow). Press [ENTER].