Transmitter M300 FLOW 31
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M300 FLOW
Printed in Switzerland 52 121 319
7 Calibration
(PATH: Cal)
The calibration key allows the user one-touch access to the calibration features of the
transmitter.
7.1 Enter Calibration Mode
While in Measurement mode press the key. Enter the security code if necessary (see section
9.4) Press the or key to select the type of calibration desired. The options are ”Sensor”,
”Meter”, or ”Analog”.
C
NOTE: To exit Calibration mode at any time press the and keys simultaneously (Escape).
The transmitter returns to the Measurement mode and the old calibration values remain active.
C
NOTE: If only Calibrate Sensor function is accessible, the Calibrate Unlock function is set to ”No”.
To access all calibration functions using the one-touch Cal key, go to Service/Calibrate menu
and change Unlock setting to ”Yes”. Refer to Section 11.2 for Meter and Analog calibration
instructions and for instructions to unlock calibration functions for Cal key. [Analog and Meter
calibration are always accessible in the Service/Calibrate menu]
C
NOTE: During calibration, the outputs will default to be held at their current values until
20 seconds after the calibration menu is exited. A flashing H appears in the upper left corner
of the display while outputs are held. Refer to Section 8.7 Hold Outputs to change the hold
output status.