Transmitter M300 FLOW 42
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M300 FLOW
Printed in Switzerland 52 121 319
8.5 Alarm
(PATH: Menu/Configuration/Alarm)
This Menu allows the configuration of an Alarm.
8.5.1 Alarm Setup
To select ”Setup Alarm”, press the or key so that ”Alarm” is flashing.
Using the and buttons, navigate to ”Use Relay #”. Using the or keys, select relay
(1, 2, 3 or 4) to be used for the Alarm and press [ENTER].
One of the following events may be alarmed:
1. Power Failure
2. Software Failure
If any of these are set to Yes then the alarm will come on and an alarm message will be
recorded if:
1. there is a power failure or power cycling
2. the software watchdog performs a reset
For 1 and 2 the alarm indicator will be turned off when the alarm message is cleared. It will
reappear if the power is constantly cycling or if the watchdog is repeatedly resetting the system.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values, selecting Yes will make the entered values the current ones
C
Note: Each Alarm Relay can be configured in either a Normal or Inverted state. In addition,
a Delay for the activation can be set. For more information, refer to Section 8.5.