7.7 ISFET calibration
h
NOTE: When measuring with an ISFET sensor, the nominal zero point must be adjusted each
time a new sensor is connected (to adjust the operating point). The adjustment for that sensor
remains stored in the transmitter. Immerse sensor in a zero point buffer (6.5 ... 7.5). Make a
mV calibration and enter for point 1 the value 00.00 mV. (see chapter 7.7.4 “mV calibration”).
Afterwards a two-point calibration (see chapter 7.7.2 “Two-point calibration”) of the ISFET sen-
sor is recommended to achieve best measuring results.
7.7.1 One-point calibration
Enter ISFET calibration mode as described in section 7.1 “Enter Calibration Mode”.
Select 1 point Calibration. With ISFET sensors a one point calibration is always performed as a
offset calibraton.
The following calibration steps are the same as described for pH sensors (see chapter 7.6.1
“One point calibration”).
7.7.2 Two-point calibration
Enter ISFET calibration mode as described in section 7.1 “Enter Calibration Mode”.
Select 2 point Calibration.
The following calibration steps are the same as described for pH sensors (see chapter 7.6.2
“Two-point calibration”).
7.7.3 Process calibration
Enter ISFET calibration mode as described in section 7.1 “Enter Calibration Mode”.