Transmitter M400 39
© 02 / 2010 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400
Printed in Switzerland 52 121 378
4.5.9 TB3 – Typical wiring for analog dissolved carbon
dioxide sensor
Cable
CO
2
electrode
Temperature
probe
Reference electrode
Sensing electrode
1
2
3
4
5
6
7
8
9
Jumper
h
NOTE: Jumper Terminal 3 and 4 has to be installed.
M400 connector:
1: Glass
2: not used
3: Reference
4: Shield / GND
5: not used
6: RTD ret / GND
7: not used
8: RTD
9: not used