Transmitter M400 12
© 02 / 2010 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400
Printed in Switzerland 52 121 378
2.3 Ex Classication
h
NOTE: The Ex classification is valid for the transmitters M400 Type 1, M400 Type 2 and
M400 Type 3. For the transmitter M400 Type 1 Cond Ind the approvals are in preparation.
Type plate
Special condition(s) X
1. The strength of the device’s casing corresponds only to the low degree of mechanical risk
and must therefore be additionally protected by suitable measures against mechanical im-
pact effect.
2. Due to the risk of electrostatic charge, the device may only be cleaned with a damp cloth.
This instruction is fitted on the device with a separate warning shield stating the following:
WARNING – CLEAN PLASTIC PARTS WITH MOIST CLOTH ONLY.
3. In accordance with guideline 94/9/EG separately certified cable and bushes in addition to
sealing plugs/caps may be used.
4. Unused openings must be closed with the sealing plugs/caps shown under point 3.
h
NOTE:
1. The transmitter M400 Type 1, M400 Type 2, M400 Type 3 is a device of device type II cate-
gory 3G in accordance with RL 94/9/EC (ATEX 95) appendix I, which may be used in accor-
dance with RL 99/32/EG (ATEX 137) in zone 2 and in the gas groups IIA, IIB and IIC, which
are explosive because of inflammable materials in the range of temperature classes T1 to T4.
During use/installation, the requirements in accordance with EN 60079-14 must be com-
plied with.
2. The permissible ambient temperature range is –10 °C to + 50 °C.