6
Installation examples (drawings) GPro™ 500 TDL Series
Installation examples
Dimensions of the
590 mm probe
607 (23.90”)
590 (23.23”)
100 (3.94”)
400 (15.75”)
90 (3.54”)
490 (19.29”)
3
4
2
1
½" NPT
G ¼"
G ¼"
50 (± 0,3 mm)
(19.68 [± 0.01"])
Definition of the lengths:
Nominal path length, the default
length when GPro 500 is delivered.
It corresponds to the effective path
length without purging.
Probe length, the physical length
of the probe.
Insertion length, part of the probe
that has to protrude into the pipe
for effective purging.
Effective path length, when con-
figuring the GPro 500 with the
M400, the double value of the
effective path length must be keyed
in (2effective path length).