RC Calibration
Basics: The F1 and F2 keys serve the following functions: F2 is used to cycle through menu items, F1
selects the menu item showing on the display.
Calibration
1. Instrument must be stable and warmed up. Weighing pan must be empty. Tare to zero. Press F1
2. At the upper right corner of the readout, either CE or CI will show. To calibrate using the internal
weight, proceed to step 3. For external calibration, proceed to step 4
3. Press F1. The display will show C as the internal calibration takes place, and returns to normal
weighing mode automatically. Don't disturb the instrument while C is showing. Internal calibration
is complete.
4. For external calibration, press F2 repeatedly until CE shows at the upper right corner of the display,
then press F1. The display will indicate the required calibration weight value. If the instrument has a
keypad AND you know the corrected value of the calibration weight, it can be entered now via the
keypad.
5 After placing the external calibration weight at the center of the weighing platform, press F1 .
6 The readout will show C during calibration, then return to normal weighing automatically.
Linearization
1. Empty weighing pan. Unit must be stable and warmed up.
2. Press TARE until menu choices appear at the upper right corner of the readout.
3. press F2 repeatedly until L I shows at the corner of the readout, then press F1.
4 Push TARE
5. The display shows C until internal linearization completes , then returns to normal weighing
automatically.