P
ower supply 100 to 240 V AC, or 20 to 30 V DC, 10 VA
Frequency for AC 50 to 60 Hz
Number of outputs (analog) 4
Current output 0/4 to 20 mA, 22 mA alarm
Measurement Error through analog outputs < ± 0.05 mA over 1 to 20 mA range
< ± 0.10 mA over 0 to 1 mA range
Analog output configuration
Linear, Bi-Linear, Logarithmic, Autoranging
L
oad max. 500
Ω
Display Backlit LCD, 4 lines
Keypad 5 Tactile feedback keys
L
anguages 5 (english, german, french, italian, and spanish)
PID process controller Pulse length, pulse frequency or analog control
Hold input Yes
Alarm contact Yes (alarm delay 0 to 999 s)
Connection terminal Detachable screw terminals
Relays – 2-SPDT mechanical rated at 250 VAC, 3 Amps,
– 2-SPST mechanical rated at 250 VAC, 3 Amps
– 2-Reed rated at 250 VAC or DC, 0.5 Amp switching
Digital Input 2
Port USB, Type B connector
Alarm Relay delay 0 – 999 s
Mains fuse 1.0A slow Blow type FC
Max. cable length ISM 80 m
Environmental specifications
Storage temperature –40 to 70 °C (–40 to 158 °F)
Ambient temperature operating range –10 to 50 °C (14 to 122 °F)
Relative humidity 0 to 95%
Emissions According to EN55011 Class A
UL Electrical Environment Installation (overvoltage) Category II
Material
ABS/polycarbonate
Hazardous areas FM: cFMus Class I Division 2 (pending)
Atex: ATEX Zone 2 (pending)
Mechanical specifications
Dimensions
Housing – (H3W3D)* 14431443116 mm
Fr
ont bezel – (H
3W)*
150
3150 mm
Max. D – panel mounted 87 mm
Weight 0.95 kg (2 lb)
Insulation
/
Rating
NEMA 4X, IP 65
* H = Height, W = Width, D = Depth