45
AM50 AM100 PM100 PM200 PM400 PM1200 PM5003 PM480 PM2500
Comparator DeltaRange DeltaRange
Readability 0.1 mg 0.1 mg 0.001 g 0.001 g 0.001 g 0.001 g 0.001 g 0.01 g 0.01 g
- Fine range (recallable) - - - - - - - 0.001 g 0.001 g
Weighing capacity 51 g 110 g 110 g 210 g 410 g 1200 g 5100 g 410 g 2100 g
- Fine range (recallable) - - - - - - - 80 g 500 g
Taring range (by subtraction) 51 g 110 g 110 g 210 g 410 g 1200 g 5100 g 410 g 2100 g
Reproducibility (s) 0.1 mg 0.1 mg 0.5 mg 0.5 mg 0.001 g 0.001 g 0.0015 g 0.003 g 0.003 g
- Fine range - - - - - - - 0.001 g 0.001 g
Linearity ± 0.2 mg ± 0.2 mg ± 0.002 g ± 0.002 g ± 0.002 g ± 0.002 g ± 0.01 g ± 0.005 g ± 0.005 g
- Fine range - - - - - - - ± 0.002 g ± 0.002 g
Sensitivity drift / °C (10 30 °C) 2 x 10
-6
2 x 10
-6
4 x 10
-6
4 x 10
-6
3 x 10
-6
1.5 x 10
-6
2 x 10
-6
4 x 10
-6
1.5 x 10
-6
Stabilization time
1)
2.5/4/6 s 2.5/4/6 s 1.5/2/3 s 1.5/2/3 s 1.5/2/3 s 2.5/4/6 s 9/12/15 s 1.5/2/3 s 2.5/4/6 s
Update speed 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s
Display
2)
FD FD FD FD FD LCD LCD FD LCD
Result deviation
in inclined position (1:1000) 0.5 mg 0.5 mg 0.005 g 0.005 g 0.005 g 0.005 g 0.02 g 0.005 g 0.005 g
Weighing pan Ø 80 mm Ø 80 mm Ø 130 mm Ø 130 mm Ø 130 mm Ø 130 mm Ø 130 mm Ø 130 mm Ø 130 mm
Calibration weight
3)
50 g/E2 100 g/E2 100 g/F1 100 g/F1 200 g/F1 1000 g/E2 5000 g/E2 100 g/F1 1000 g/F1
Net weight 6.0 kg 6.6 kg 3.8 kg 3.8 kg 3.8kg 3.8 kg 6.6 kg 3.8 kg 3.8 kg
Balance housing (W x D x H) in mm 194 x 316 x 252 194 x 316 x 337 194 x 316 x 68 194 x 316 x 337 194 x 316 x 68
Power consumption 6 VA
Fusing 63 mA/220 V
125 mA/110 V
Technical data
Technical data for individual models